| Summary: | A 37-year-old female underwent an emergent implant of a covered self-expandable stent graft (8mm x 10cm Viabahn, W. L. Gore) due to acute bleeding from a femoral access following closure of a patent foramen ovale. Over the following months, intermittent claudication developed, progressively more incapacitant. Work-up revealed near-occlusion of the stent graft due to edge stenosis. Elective explant of the stent graft was performed, revealing extensive proximal and distal edge miointimal hyperplasia (marked with *). Endarterectomy was performed and the arterial defect was corrected with a bovine pericardium patch. |
|---|---|
| Country: | Portugal |
| Document type: | journal article |
| Access type: | Restricted |
| Associated institution: | Angiologia e Cirurgia Vascular |
| Language: | English |
| Origin: | Angiologia e Cirurgia Vascular |
| Summary: | A 37-year-old female underwent an emergent implant of a covered self-expandable stent graft (8mm x 10cm Viabahn, W. L. Gore) due to acute bleeding from a femoral access following closure of a patent foramen ovale. Over the following months, intermittent claudication developed, progressively more incapacitant. Work-up revealed near-occlusion of the stent graft due to edge stenosis. Elective explant of the stent graft was performed, revealing extensive proximal and distal edge miointimal hyperplasia (marked with *). Endarterectomy was performed and the arterial defect was corrected with a bovine pericardium patch. |
|---|